She is not getting time spectra using RTC option. The AUX input was not used. Is it not possible to get time spectra with RTC option without using AUX input? She opted for software start but no spectra came.
We are trying to communicate with a NHQ High Voltage unit via its RS232 port. We send (for instance) "U1 <CR><LF>" to determine the voltage on channel A. However, we are sent back (in ASCII) "U1<LF>????<CR><LF>". Many times we get a timeout error.
It's possible to control the MPA server by sending commands over a TCP/IP net ? - what are the messages to send : can we do all the functions available in the DLL?
Is it possible to inspect MPA-3 datafiles under Linux (offline, without connected hardware) (MPANT seems not to run under wine)?
Would it be possible to run more than two of your P7888 Cards on a single PC in order to have more than 8 input channels? Is the total transfer rate for ALL boards (summed > together) still (about) 20 MHz?
We wanted to control the NHQ105M unit through RS232 using computer. We have tried to do the same using HyperTerminal of windows. We were able to get the echo but all the commands in the table of the manual did not work?
We urgently need to control this NHQ205M unit through computer interface through the given RS232 port. But the corresponding manual has misplaced some where and now we dont know the compand instructions codes for its control?
Please could you let me know if there are any known incompatibilities between P7889 and this DELL workstation I intend to buy? Please could you also let me know if Windows XP 64bit is supported by your drivers, or wether I will need to install a 32bit OS?
Is there anyway to save the P7887 data from each scan using the calibration variable (say, time of flight) instead of the bin number?
Is there in fact any more documentation on using the DDE method, in addition to what I have referenced below in the manual?
When using the logsums command we noticed that the logfile does not update until we exit out of the program or the logfile name is changed. Is there any way to get the sums to automatically be saved into a .txt file after we have replayed the data?
Can P7888 be controled by Matlab?
We want to use an external TTL trigger signal to start or stop or clear an acquisition, thus allowing to record sequential spectra. What would be the best way to achieve this? What signal should we use (TRIGGER? GO Line? Other I/O?).
The measurement is stopped with acquisition times of multiples of 2 seconds when using my P7888-1E and time preset mode using DMA. Where does it come from?
How can we access via DLL the data stream of P7889 count events ?
Is it possible to read "on line" all the events with an external application written in VB.NET or Labview or C++ ... ?
Is it possible to do time-of-flight measurements using MPA-3 with RTC by acquiring an ADC spectra for a time window and starting the time by a chopper signal?
I use fast position-dependent detectors with delay-line read out. The maximum count rate that I can reach using two 7072T TDC's in coincidence and MPA-3 is only about 350 kHz per input. Is there a solution that can handle a higher count rate?
Is it possible to have two P7888-2 cards working as if it were a four-channel one?
On what conditions in the ADC/ MPA-3 can result in counts (especially large numbers of counts) appearing in MPA-3 channel 0. The customer is experiencing this. How can he eliminate it?
When using your PSD module, 2160A, Do I have to use other FAST ComTec modules, 2128 as a CFD ? I am not sure your PSD, 2160A can be used with other NIM module like the Ortec CFDD, 583B, and Canberra CFT SCA, 2035A?
My 7033 NIM bin cannot supply the specified current at 24 V when used with a special NIM module. What can be the reason?
Waveform Digitizers / Transient Recorders - No frequently asked question till yet.
X-Ray and Gamma Detector Systems - No frequently asked question till yet.
Please advise how can we check PAS-2. My customer gets no counts. He checked PAS-2 signal on oscilloscope but is not seeing any pulses. There is a possibility that either source is dead or Detector problem or PAS-2 malfunction.
We would like to remote control the HV supply via RS232. In the manual it mentions that there already exists a LabView VI. We would be interested in using this VI and were wondering whether we could be sent a copy, or details of the code used?
Nuclear Pulse Generators - No frequently asked question till yet.
Pulse & Delay Generators - No frequently asked question till yet.
Is there for the Ultra Fast Multiscalers the FAST ComTec three-year-warranty included?
About the 4K and 16K FIFO: Can you give me some firm ideas on how to recommend to customers which they need?
How can I get in contact to FAST ComTec?
How can I find the right product for my project?