Model P7882 |
PCI based, 350 MHz Dual Input Multiscaler/TOF/Photon Counter
FEATURES
Using one input: 100 ns to 50s, manual and external
Using two inputs: 200 ns to 25s, man. and external, programmable in steps of 25ns
1 Input: 256... 512k
2 Inputs: 256 ... 256k
Start, Count 1, Count 2, Ext. channel advance
optional 0...10 V, 12bit programmable voltage or X-ramp output with sawtooth waveform and programmable start value;
+/-2.5V or 0...10V (jumper selectable), 12 bit
APPLICATIONS
DESCRIPTION
The P7882 is an advanced dual-input Multiscaler / TOF / Photon Counter designed for fast sweep repetition rates and counting rates in excess of 350 MHz (400 MHz typical). The minimum dwell time (time-bin width) is 100 ns, two inputs accept counting rates of >350 MHz each. Both inputs and the START input are equipped with computer adjustable discriminators.
A special design prevents lost counts and double counting of input pulses.
The very large data memory has a 32 bit storage capacity and can be segmented to enable to accumulate successive measurements.
Spectra can be displayed in a two-dimensional array. This can be used to display successive sweeps to make - for example regression studies where even small changes are very visible.
Sweeps can be summed for realtime signal averaging - this means that as a new sweep is acquired it will be added to the sum of previous sweeps.
For the use in remote places an autostore function can be selected that will store spectra and setup parameters at selectable intervalls.
The P7882 can be fully remote controlled by a host computer. "GO"-line compatibility enables the P7882 to start and stop accumulation synchronously with other FAST ComTec products like the MS-12 Timer/Scaler, the Multi Paramer System etc.
Extensive Operating Software is based on the MCDWIN software - a proven - easy to use graphical user interface for mouse and keyboard operation. High resolution display with linear and log scales, autoscaling, ROI, zoom, Gaussian-Fit, FWHM calculation are just a few features MCDWIN offers. Batch processing with MACRO's is a flexible way to allow the user to execute task oriented batch programs.
The interactive display software can be set to display spectra with a high resolution, linear- or log scale. Screen dumps allow text insertion and graphic editing with a variety of tools. This package also supports all graphic output devices installed under WINDOWS 2000 and XP.
A unique feature of the P7882 is the selectable autostore facility. In case of a power failure theP7882 will lose just the last minute of data. As all setup parameters are stored as well, a measurement can be continued as soon as power has returned.
SPECIFICATIONS
Dwell Time Modes: programmable internal-, external- and manual channel advance
Dwell Time per channel: 100 ns (200 ns for two inputs) to 50 s, progr. in steps of 25 ns, and external
Countrate: 350 MHz, typically >400 MHz
Deadtime between time bins: zero
Bin accuracy (leading/trailing edge) rel. to the Reference Clock +/-0.15ns, typical +/-0.07ns
End-of-sweep deadtime: zero
Trigger Delay: 25ns +/-12.5ns
Abort: ends the sweep at the end of the actual time bin
Trigger / Start uncertainty: < 25 ns
Sweep Preset Range: 1 to 232
INPUTS:
START/TRIGGER: +/-5V range, 50 W/1 kW slope and threshold programmable in steps of 2.44mV,
COUNT1 and COUNT2: +/-5V, 50 W/1 kW , falling edge sensitive, threshold programmable in steps of 2.44 mV, count rate capability 350 MHz, typically >400 MHz
EXTERNAL CLOCK: 50 kW, rising edge sensitive
CONNECTORS:
Start/Trigger: LEMO 00 type connector
Count1, Count2: LEMO 00 type connector
Clock/Abort: LEMO 00 type connector
Sync-Out: LEMO 00 type connector
D-SUB 15 pin female connector:
Mossbauer Control lines
Analog Ground
Digital Ground
+5V OUT
DAC OUT: 0...10V
DIG I/O: 8-bit, very versatile
Power Requirements:
+5V / 1400mA / 7W typ. (1600mA / 8W max.)