Model TA1000B Fast Pulse/Timing Preamplifier



FEATURES

Small signal bandwidth DC … 1GHz (x50 model)

Voltage gain 20dB (x10), 34dB (x50), 40dB (x100) and 46dB (x200)

DC coupling

Closed loop OP-Amp design

Very low noise

High output drive

Single supply operation / internally generated

bipolar supply / internal supply regulation

Bandwidth limited (BWL) option available for

further improved noise performance

APPLICATIONS

Pre-amp for ultra fast detectors (MCP, PMT, …)

Oscilloscope and transient recorder pre-amp

Photon-/Ion- counting

Wideband signal processing

 


DESCRIPTION

The TA1000B-x models are fast, very low noise pulse pre-amplifiers with a small signal bandwidth of 400MHz … 1.0 GHz depending on the model.

Each model is av ailable with a bandwidth limited (BWL) option which f urther reduces the noise floor.

A unique feature for such high speed amplifiers is the DC coupling. DC coupling avoids count rate effects due to non DC balanced pulse trains and the corresponding charging of coupling capacitors.



Simulated Voltage Gain / Simulated Pulse Response



(*) 200mVeff output signal into 50Ω load
(*) Signals are voltage shifted for better comparison (**) Input signal to each amplifier is scaled according the gain



Pulse Response









TA1000B-10 (x10), Input 20mV/div, Output 100mV/div
TA1000B-10BWL (x10), Input 20mV/div, Output 100mV/div
TA1000B-50 (x50), Input 5mV/div, Output 100mV/div
TA1000B-50BWL (x50), Input 5mV/div, Output 100mV/div
TA1000B-100 (x100), Input 2mV/div, Output 100mV/div
TA1000B-100BWL (x100), Input 2mV/div, Output 100mV/div
TA1000B-200 (x200), Input 1mV/div, Output 100mV/div
TA1000B-200BWL (x200), Input 1mV/div, Output 100mV/div


In the following scope pictures you see the pulse response for negative output signals starting at 0V and falling down to –400mV. The input pulse amplitudes are selected according the gain of each amplifier.


The lower window of each plot shows details of the corresponding signals in the upper window. There is also a (redcolored) histogram of the output signal jitter at a-100mV or –130mV threshold. The jitter's Peak-to-Peak value is visible at "PkPk" and its standard deviation in the "RMSΔ" readout.

This jitter histogram gives a good indication of the timing accuracy and resolution that can be expected. And, one can very well see that the optimum threshold setting for timing measurements is often not at half of the signal's amplitude but at some other level not too far from idle voltage where the slew rate is at maximum.


Max. Output Noise Voltage

Normally the noise is given input referred, so to speak, it can be compared to the source signal levels. For

timing applications it is often more depicting to plot the total output noise of an amplifier.

In the following scope pictures the output noise voltage of our TAx-amplifiers is accumulated over 10,000

waveforms corresponding to about 40 minutes of measurement time. Used was a TEK11801C digital sampling

scope with a 12.5GHz sampling head. Thus, the displayed noise voltage is accumulated over a long period and also over the full bandwidth of each amplifier. The TAx's inputs were shortened, i.e. ZSource = 0Ω.

 












Sampling Head alone, 2mV/div
TA1000B-10 BWL (x10), 5mV/div
TA1000B-10 (x10), 5mV/div
TA1000B-50 BWL (x50), 10mV/div
TA1000B-50 (x50), 10mV/div
TA1000B-100 BWL (x100), 20mV/div
TA1000B-100 (x100), 20mV/div
TA1000B-200 BWL ( x200), 20mV/div
TA1000B-200 ( x200), 20mV/div


On the right side of each plot you can see a (red colored) histogram of all the voltage samples in the respective picture. This gives the probability distribution of the noise voltage levels. And, you can find some analysis data on the respective voltage distribution: Mean = average value, RMSΔ = σ = standard deviation, PkPk = Peak-to- Peak voltage = max. – min. sample voltage, μ ± 1σ = percentage of samples that fall within ±1 standard deviation of the mean (.±2σ, ±3σ respectively).