INPUTS (front panel)
Signal Inputs: BNC-Connector accepts +25mV to +10V (standard / +8V optional) linear pulses, Zin = 1k Ohm, risetime: ≥ 100ns to 100µs, fall time: 200ns to 100µs (optimized for gaussian shaped pulseswith shaping times from 250 ns to 25µs, dc coupled or a varying voltage level when used in SVA mode.
Gate/Start inputs: BNC-Connector accepts positive TTL for the Gate mode
OUTPUTS (front panel)
SCA: TTL output in PHA and SVA mode (TTL pulse of approx. 0.5µs duration. One output pulse for each input signal that falls between the ULD and LLD setting).
Outputs monitor levels in adjustment mode
OUTPUTS (rear panel)
Data: 13 bit binary data lines and data transfer commands, 25-pin female D-Sub-connector.
General
Temperature range: 0oC to +50oC storing
Power Requirements
+24V, 100mA, -24V, 90 mA, +12V, 150mA, -12V, 200 mA, + 6V, 700mA, -6V, 430 mA
Physical
Size: single width NIM module (1.35 x 8.71 inches;
3.43 x 22.13 cm) as per TID - 20893
Shipping weight: 1.8 kg (net 1.0 kg)
|
Front Panel Control push-buttons
Conversion Range: Push button switch selects quantization of 256, 512, 1k, 2k, 4k, or 8k channels for full scale input.
Coincidence/Anticoincidence: Push button switch selects COINCIDENCE/ANTICOINCIDENCE mode
LLD: adjustment by up/down pushbutton sets the lower level discriminator. The voltage is displayed on the DEADTIME/LEVEL LED indicator and output on the monitor connector for precise adjustment
ULD: adjustment by up/down pushbutton sets the upper level discriminator. The voltage is displayed on the DEADTIME/LEVEL LED indicator and output on the monitor connector for precise adjustment
Front Panel Mode Control Switches
PHA: Pulse Height Analysis mode.
SVA: Analog Sampled Voltage Analysis
Indicators
DEADTIME/LEVEL: Dual 10 LED-bar indicates activity of the ADC. Range 0 to 100% deadtime or actual voltage level indication of ULD and LLD setting
RANGE/MODE: ADC range 256 to 8k in binary steps
|