INPUTS
INPUT: -5mV to about -3.5V linear pulses in the BLD part (ref. common mode range of ultra fast comparators) and -10mV to -2.5V linear pulses in the SCA part, risetime > 700ps typ., Rin = 50W, dc coupled, front panel BNC connector, minimum input pulse width > 1ns (BLD part)
DELAY: 2 BNC connectors for an external delay cable to form the internal constant fraction signal (for specific length see section Performance).
OUTPUTS
INSPECT: Displays output signal of zero crossing discriminator for use in adjusting the time walk.
BDL neg.: negative current output, providing -32mA into 50W, risetime approx. 2ns, pulse width > 5ns nominal, dc coupled
BLD pos.: positive voltage output providing 1.5V (min.) into 50W, risetime approx. 4ns, width (in normal deadtime mode) adjustable by front panel trimming potentiometer, the width then determines the internal deadtime
SCA neg.: negative current output, providing -32mA into 50W, risetime approx. 2ns , pulse width > 5ns nominal, dc coupled
SCA pos.: positive voltage output, providing 2V (min.) into 50W, risetime approx. 4ns, width as BLD pos.
ULD: positive voltage output, providing 1.5V (min.) into 50W, risetime approx. 4ns, width as BLD pos.
|
CONTROLS
THRESHOLD: Front panel 10-turn locking dial potentiometer to set acceptance threshold for input pulses: range - 5 mV to - 1.2V.
LLD: front panel 10 turn locking dial potentiometer to set lower discriminator threshold for input pulses in the SCA part (range approx. -10mV to -2.5V)
ULD: front panel 10 turn locking dial potentiometer to set upper discriminator threshold for input pulses in the SCA part (mode ULD on, range approx. -10mV to -2.5 V)
Z/C adjust: (walk adjust) front panel trimming (screwdriver) potentiometer to compensate walk of the internal zero crossing discriminator
CF-SR: front panel two position switch to select the modes of timing
ULD on/off: selects window or integral mode of operation in the SCA part
DT n/up: selects mode of deadtime operations: in n (= normal) position the deadtime is given by the width of the positive output signals (BLD, SCA or ULD). Width is adjustable from 10 ns to 100 ns by a front panel (screwdriver) potentiometer (up = updating)
INDICATORS
Rate: LED indicates activity of the discriminator. Color of the LED changes by count rate of the BLD part: green up to approx. 5kHz, orange from approx. 5kHz to approx. 10kHz, red above (for statistical input pulses)
|