FEATURES
- 100 MHz Count Rate
- 1000: 1 Dynamic Range
- Five Operating Modes
- Walk < 30 ps typically for 100: 1 Range
- Simultaneous Dual Positive and Negative Outputs
- Multicolor Count Rate Indicator
- DC coupling
- Easy to setup and use
DESCRIPTION
The Model 2128A is a fully dc-coupled constant fraction discriminator with a dynamic range of up to 1000: 1. Three operating modes provide optimum time resolution for many detector types and applications. For example:
CFT-Constant fraction mode for fast detectors
CFRR-Slow rise time reject mode for Ge detectors
LET-Leading edge mode for single photon counting
In addition, internallyselectable program modules provide for
1) User change of fraction from the standard 40%,
2) Operation of the unit as a Zero Crossing Discriminator for bipolar inputs, and
3) Operation of the unit as a Leading Edge Discriminator without termination of the delay ports.
The Model 2128A has a dc-coupled 50 ohm input which accepts negative pulses. The constant fraction composite signal is formed by the algebraic sum of a direct, attenuated signal path and a delayed. unattenuated path. The delay time is user selected by cable delay. Optimum selection of this external delay provides full compensation for timing distortions due to both amplitude- and rise time variations in the input signal.
Four simultaneous, independent output signals are provided. The two positive outputs are adjustable in width, the width duration sets the internal dead time required to suppress spurious outputs due to input signal anomalies. The two negative outputs are fixed-width pulses keyed to the start of the dead time period.
The front panel WALK ADJUST control and INSPECT OUTPUT permit the user to trim the time walk characteristics of the experimental setup for optimum timing resolution. A novel frontpanel LED indicates count rate by color change.
|