General
Description: The Model 2128 is a fully dc-coupled constant fraction discriminator with a dynamic range of up to 1000: 1. Three operating modes provide optimum time resolution for many detector types and applications.
For example:
In addition, internallyselectable program modules provide for 1. User change of fraction from the standard 40%, 2. Operation of the unit as a Zero Crossing Discriminator for bipolar inputs, and 3. Operation of the unit as a Leading Edge Discriminator without termination of the delay ports.
The Model 2128A has a dc-coupled 50 ohm input which accepts negative pulses. The constant fraction composite signal is formed by the algebraic sum of a direct, attenuated signal path and a delayed. unattenuated path. The delay time is user selected by cable delay. Optimum selection of this external delay provides full compensation for timing distortions due to both amplitude- and rise time variations in the input signal. Four simultaneous, independent output signals are provided. The two positive outputs are adjustable in width, the width duration sets the internal dead time required to suppress spurious outputs due to input signal anomalies. The two negative outputs are fixed-width pulses keyed to the start of the dead time period. The front panel WALK ADJUST control and INSPECT OUTPUT permit the user to trim the time walk characteristics of the experimental setup for optimum timing resolution. A novel frontpanel LED indicates count rate by color change.
Features:
SpecificationInputs input: Accepts - 5 mV to - 5 V linear pulses: width:> 1ns, Zin= 50 Ohms, dc coupled; front panel BNC connector. Performance
Dynamic Range: 1000 : 1 CF Mode Walk: <±50ps (typically ±30ps) for -30mV to -3V range with <2 nsec rise time. Counting Rate: to 100 MHz, limited by dead time (OUTPUT WIDTH setting). Pulse Pair Resolution: <10 ns, or as limited by dead time. Threshold Stability: Better than ±0.02 %/oC (± 200 ppm/oC) Temperature Range: 0 to +50oC Threshold Linearity: ±0.25% Integral Typical Cable Lenghts: (RG-58) For Plastic, Nal and Si (S.B.) detectors: 0.5 to 1.0 m For Planar Germanium detectors: 1.0 to 2.0 m For Coaxial Ge: 2.0 to 4.0 m Typical Power Requirements: Standard version +6 V - 150 mA, +12 V - 70 mA - 6 V - 450 mA, -12 V - 100 mA 12V Version: +12 V - 220 mA* -12 V - 550 mA* Pysical size: Single width NIM module 3,43 X 22,12 cm (1.35 X 8.71 inches) per TID-20893 (rev.) Net weight: 0.9 kg (2.0 lbs.) Shipping weight: 2.2 kg (4.9 lbs.) Accessories included: Fraction module f = 0.4 Leading edge module Zero cross module Special version using ±12 V available on special request.
* This power exceeds the normal bin allotment of 167 mA for a singlewidth module for the 12V version
|
||||||||||||||||||||||||||||||||