Description
The Model 7072T is a unique dual channel device that can operate either as an ultra fast analog-to-digital converter with a fixed conversion time of 500 ns*, a time-to-digital converter (TDC) or an analog SVA . The 7072T is ideally suited for high counting rate applications in energy and time Spectroscopy. The combined ADC/TDC function finds applications in time resolved photon, ion-, neutron- and X-ray imaging.
The built-in Single Channel Analyzer has a separate output on the front panel. This output can be used in the ADC and SVA mode. An analog output is available for monitoring the TDC.
In the SVA mode a voltage level or waveform can be sampled when a gate signal is applied.
*) Conversion time of ADC. For dead time calculation add time-to-peak and 100 ns transfer time to data buffer.
Features
- Three individually selectable modes of operation for each input channel:
- ADC mode with 500 ns Fixed Conversion Time
- TDC mode with <<25 ps Time Resolution
- ADC/SVA (Analog Sampled Voltage Analysis) mode
- Full 8k Conversion Gain and Range
- Dual 10-LED Indicator bar for Instantaneous Dead-Time display, level setting
- Lower Level and Upper Level Discriminator on each ADC input
- Three modes of operation independently selectable for each input channel: Pulse Height Analysis with automatic peak detection / Time-to-Digital conversion and Analog Input Sampling selectable by frontpanel switch
- Simultaneous Single-Channel Output / TDC Monitor
- Designed for easy interfacing to our MCA-3, SPA-3, MPA-3, MPA4, MPA4T multichannel analyzers, MCA's from other manufacturers (special cable and external ADC interface required)
Specifications
|
INPUTS (front panel) |
Front Panel Control push-buttons |
ADC Performance
| ADC Performance | Conv. Range 4K | Conv. Range 8K |
| Resolution: 500 ns flattop pulses | <3.4 chan FWHM | |
| Integral Nonlinearity linear fit | +/-0,16% typical | +/-0,16% typical |
| Differential Nonlinearity incl. Effects of int.NL | +/-0,7% | +/-1,1% |
| Differential Nonlinearity for 99% of useable channels | +/-0,5% | +/-0,7% |
| Counting Rate Capability | > 1.000.000 events/s | |
| Peak shift up to 700 kcounts/s SVA mode | +/-0,006% | |
| Peak shift up to 1.000 kcounts/s SVA mode | +/-0,08% | |
| Unused channels | approx. Top 1% of range | |
|
PERFORMANCE ADC Conversion Time: 500 ns fixed conversion time ADC Deadtime: time-to-peak + 500 ns fixed conversion time + data transfer time of 100 ns to buffer (there could be an additional dead time if the MCA can not accept the data rates generated by the 7072T - this depends on MCA used) Typical ADC deadtime for fast rising input pulses is less than 2 µs per event using FAST ComTec MCA's Gain stability: better than 50 ppm/oC Baseline stability: 50µV/oC | ||
TDC Performance
| TDC Performance | Conversion Range 8K | |
| Resolution @ 50 ns time range | < 4.1 chan FWHM | |
| Integral Nonlinearity 50 ns range, linear fit | +/-0,12% typical | |
| Differential Nonlinearity 50 ns range | +/-0,5% | |
| Differential Nonlinearity for 99% of useable channels | +/-0,4% | |
| Counting Rate Stability up to 100 kcounts/s | +/-0,003% | |
|
PERFORMANCE TDC Time range: 50 ns, 100 ns, 200 ns, 500 ns, 1 µs, 2 µs, 5 µs, 10 µs, 20 µs Useable times: >12..15 ns to approx. 95% of selected range Conversion range: 256, 512, 1k, 2k, 4k, or 8k |
||